Model RX-7000i Cat.No. 3279
Measurement Range Refractive index (nD) : 1.32422 to 1.70000
Brix : 0.00 to 100.00% 
( 5 to 75°C ATC )
Minimum indication Refractive index (nD) :
0.00001 (0.0001)*
Brix : 0.01% (0.1%)*
Temperature: 0.01°C 
*Factory default setting
Measurement Accuracy
*Repeatability
Refractive index (nD) :
±0.0001
*±0.00005
Brix : ±0.1% 
*±0.02%
[*1]
Temperature control range 5.00 to 75.00°C
(No lower than 10°C below the ambient temperature and no higher than 55°C above the ambient temperature)
Power Supply AC100V to 240V, 50/60Hz
Power consumption 90VA Dimensions & weight 37×26×14cm, 7.0kg
(Main Unit only)
Mode MODE-1:
Displays the measurement value once the sample reaches the target temperature. 

MODE-2:
Measures refractive index and temperature at fixed intervals and displays the estimated measurement value at the target temperature. 

MODE-3:
The thermo-module can be turn off. Without temperature control, the measurement value is displayed in 4 seconds after the START key is pressed.

MODE-S:
Displays the measurement value once a certain level of sample stability is achieved.
Output terminals • Printer (for ATAGO digital printers)
• Computer -USB
Warranty 2 years standard (3 years with product registration) excluding touchscreen
Optional • Digital Printer DP-63 : Cat.No.3118
• Digital Printer DP-AD : Cat.No.3123
• MAGIC™(metal) : RE-56180
• MAGIC™ (resin) : RE-56185
 Flow cell attachmen: RE-56173
• 10% Sucrose (±0.01%) : RE-111001
• 20% Sucrose (±0.01%) : RE-112001
 30% Sucrose (±0.01%) : RE-113001
 40% Sucrose (±0.02%) : RE-114002
 50% Sucrose (±0.02%) : RE-115002
* Calibration Certificate : NIST-traceable certificate. This will increase the lead time of the order by 2 to 3 weeks. Contact an ATAGO representative for details.
Parts  Printer paper (regular) for DP-63 : RE-8412
• Printer paper (long-term storage)for DP-63 : RE-8414
• Printer paper for DP-AD : RE-89403 
• Ribbon cassette for DP-AD : RE-89402
• Fan Filter Replacement : RE-58001
[*1] When measuring a standard sucrose solution of up to 50% Brix or standard refractive index solution in MODE-1 at 20°C

 